Small Angle X-Ray Scattering

Beamline X33 is a bending magnet beamline for biological small-angle scattering on the DORIS Fan D, located in the HASYLAB Hall 3. It uses a fixed wavelength of 1.5 A. The endstation is equipped with a photon counting Pilatus 1M-W pixel detector (67 x 420 mm2) and an additional Pilatus 300K-W pixel detector (33.5 x 254 mm2). The beamline is optimised for low background data collection from macromolecular solutions. A two-detector setup is used to simultaneously collect small and wide angle scattering (SAXS/WAXS) patterns.

Apply for beam time

Contacts

Scientists in charge

Dmitri Svergun

Tel: +49 40 89902-125
E-mail: svergun@embl-hamburg.de

Manfred Roessle

Tel: +49 40 89902-192
E-mail: manfred.roessle@embl-hamburg.de

Small angle X-ray scattering beamline X33

Tel: +49 40 89902-153

Experimental possibilities

  • SAXS/WAXS experiments on macromolecules ranging from a few kDa to GDa
  • Time resolved measurements and temperature scans
  • Automated liquid handling sample changer
  • Sample holders for solutions, gels, films, powders
  • Resolution range: 700 to 10 A (SAXS); 17 to 4 A (WAXS)
  • Typical exposure times: 15-120 s

Facts

Synchrotron source   DORIS bending magnet
Monochromator   Asymmetric cut 7° crystal Si (111), horizontally focusing
Mirror   Rh coated, Zerodur, vertically focusing
Acquisition system   CANbus, TINE
Wavelength   1.5 A
Focal spot   2 mm x 0.6 mm
Flux   5 x 1011 ph/s/100mA
Slits   4 pairs, tungsten
Small automated sample changer   8 samples (200 ?l)
3 buffers (1.5 ml)
temperature controlled storage (5-45°C)
Large automated sample changer   2 x 96 samples (200 ?l)
2 x 24 buffers (1.5 ml)
temperature controlled storage (10-40°C)
Vacuum cell   30 ?l, polycarbonate windows, automated filling and rinsing, temperature controlled (5-50°C)
In-air cell   100 ?l, mica windows, manual filling and rinsing – for viscous and solid samples
temperature controlled (0-80°C)
Detectors   Photon counting Pilatus 1M-W pixel detector (67 x 420 mm2);
Photon counting Pilatus 300K-W pixel detector (33.5 x 254 mm2);
Mar345 image plate (345 mm2)
Sample-detector distance   Standard: 2.7 m (SAXS), 0.75 m (WAXS)

Recent changes, upgrades and developments

Roessle, M.W., Klaering, R., Ristau, U., Robrahn, B. Jahn, D., Gehrmann, T., Konarev, P. V., Round, A., Fiedler, S., Hermes, S. & Svergun, D. I. (2007). Upgrade of the Small Angle X-ray scattering Beamline X33 at the European Molecular Biology Laboratory, Hamburg. J. Appl. Crystallogr., 40, s190-s194.

Round, A.R., Franke, D., Moritz, S., Huchler, R., Fritsche, M., Malthan, D., Klaering, R., Svergun, D.I. & Roessle, M. (2008) Automated sample-changing robot for solution scattering experiments at the EMBL Hamburg SAXS station X33. J. Appl. Crystallogr., 41, 913-917.