Technical characteristics of beamline P12

Beamline P12 uses a high brilliance X-ray beam with low scattering background and energy tunability. The beamline is optimized for solution scattering experiments and allows for kinetic SAXS with millisecond time resolution and anomalous SAXS experiments on biologically relevant metals, such as calcium or molybdenum.

Source Petra III U29 undulator
Monochromator Double crystal Si (111)
Wavelength 0.1-0.3 nm, usually 0.124 nm
Flux 1013 ph/s
Slits Individually motorized blades of Hybrid Scatterless in a module from Xenocs
Beam size at the detector 0.2 × 0.05 mm2
Exposure unit 1.7 or 0.9 mm capillary, temperature range: 5°C to 60°C
Detectors Dectris 2D photon counting detectors:
  • Pilatus 6M
  • Eiger 4M
Sample-detector distances
  • 1.5 m (position 0)
  • 2.0 m (position 1)
  • 3.0 m (position 2, most common)
  • 4.0 m (position 3)
  • 6.0 m (position 4)
Resolution 10000 Å - 8 Å
Sample changer Arinax BioSAXS (2018 model)